Check all of the following:
That the aluminium Helium dewars in the storage area are safe (i.e. no gas escaping and making a hissing noise). The helium storage area is shown in Figure 3.1. Doors to this storage area must remain closed during scanning.
If there is gas, please contact reception, who will contact a member of YNiC staff.
If it is the O2 alarm, Do Not Enter MEG
Leave immediately and notify reception, who will immediately alert a member of YNiC staff
If it is the helium lid alarm (intermittent beeping from the DAS), you don't need to leave the MEG Scan Suite.
Contact reception, who will contact a member of YNiC staff. The staff member will remove and replace the lid on the dewar.
Check O2 level on blue box inside stimulus delivery room. The O2 level should be about 20 %. The "metal" ball should be floating in the column.
If it has fallen to a dangerously low-level (i.e. below 18 %), the alarm should be sounding. If the level is below 18 % and the the alarm isn't sounding, leave MEG immediately and notify reception, who will contact a member of YNiC staff.